Page 108 - Eagle Scientific Ltd - Catalogue
P. 108

ES
        Chromatography



         HPLC SYSTEM

         Consists of one each UV/Vis diode ray detector, analytical pump, gradient box,
         column oven, analytical/semi-preparative injection valve with sample loop,
         software, column and set of accessories.
         The analytical UV-VIS diode array detector measures at eight wavelengths
         simultaneously or full scan with speed up to 100Hz. This unit is used in liquid
         chromatography to verify analysed samples. High standard deuterium discharge
         lamp in a special socket enables easy exchange from front panel. Also includes
         halogen lamp. Detector can be controlled by Clarity or ECOMAC software.
         Three selectable flow cells with optical path length 2.4, 5 and 10mm may be used.
         The 5mm cell is included as component of the detector. Other cells are available
         on request. Clarity PDA module support for 3D measurement.
         Technical Data
         Wavelength range:                    200 - 800nm
         Spectral half-width:                   10nm
         Number of CCD elements:       512 mathematically enhanced to 4,096 by algorithm
         Accuracy of adjustment:           ±1nm
         Reproducibility:                          ±0.5nm
         Noise (Test cell, 254nm, TC 1s, 10Hz):  ± 5·10-6AU
         Drift (Test cell, 254nm):            1·10-4 AU/hod
         Time constant:                           10 - 10000ms
         Sampling speed:                         up to 100Hz
         Output:                                        digital: 1V/AU; analogue: 4 x configurable   C26-3000 part - UV/Vis Diode Ray Detector
         Communication:                        Ethernet (LAN), USB, RS232
         Display, keypad:                         VFD 140x32 pixels, 10 pushbuttons
         Power supply:                             100 - 240V 50/60Hz 110V
         The analytical HPLC pump works as isocratic pump and together with gradient box
         with degasser also as gradient pump. It is suited for analytical applications in liquid
         chromatography. It is designed with main and auxiliary pump heads connected in
         series with a piston diameter of 1/8”.
         This unit allows achieving a precise low-pressure gradient, i.e. gradual changes of
         solvent composition by mixing up to four liquids at the pump’s entry. It is also
         possible to predefine percentage of composition when using the pump in isocratic
         mode. The gradient profile can be defined from a computer. The pump is supported
         by ECOMAC and Clarity software.
         Technical Data
         Flow-rate:                                                  0.01 - 10.00ml/min
         Pumping system:                                     Two plungers dia. 1/8" connected in series
         Maximum operating pressure:              40MPa (5800 PSI, 400 bar)
         Accuracy of flow-rate (1ml/min, 12MPa, H2O):  ± 2%
         Repeatability of flow-rate (1ml/min, 12MPa, H2O):  ± 0.5%
         Accuracy of pressure measurement:   ± 2%
         Adjustable pressure limit:                      lower 0.0 - 39.0MPa; upper 1.0 - 40.0MPa
         Number of valves:                                   4 (A, B, C, D)
         Communication:                                      RS232, Ethernet (LAN), USB
         Display, keypad:                                       VFD 140x32 pixels, 10 pushbuttons
         Power supply:                                           100-240V 50/60Hz 100VA
         The Peltier heating/cooling column oven with a big compartment room can
         accommodate up to 3 columns with a length up to 25cm. Unit is controlled by   C26-3000 part - HPLC Pump
         keypad, RS232 or Ethernet(LAN) interface. Precise temperature control enhances
         separation and reproducibility and improves the quality of analysis. Usage of Peltier
         technology offers the possibility of heating and cooling, providing complete
         independence from the room temperature. With its built-in leak alarm it is a safe,
         accurate and flexible choice.
         Technical Data
         Operating temperature range:              0 - 80°C
         Temperature stability:                            ± 0.2°C
         Temperature reproducibility:                ± 0.2°C
         Temperature accuracy:                           ± 2°C
         Communication:                                      RS232, Ethernet (LAN)
         Display, keypad:                                       VFD 140x32 pixels, 10 pushbuttons
         Power supply:                                           100-240V 50/60Hz 250VA