ES
Atomic Absorption Spectrophotometer
C10-9523 2 Port PCI Serial Card (for use with C109518)
C10-9524 Deuterium Lamp
C10-9525 Photomultiplier Tube
1, 2 & 5 years operating spares available upon request. Please enquire for more
details.
For a full range of Hollow Cathode Lamps and Standards please see page 10.
ATOMIC ABSORPTION SPECTROPHOTOMETER
(AA500FPC/AA500GPC/AA500FGPC)
Atomic Absorption spectrophotometer, fully automated system. Available in a
choice of three configurations; Flame only, Graphite Furnace or Combined Flame
and Graphite Furnace. The change over from one technique to another is via a
simple key stroke within the software. All configurations have a built-in PC system
complete with pre-loaded Microsoft operating system and AAWinPro Software. The
Flame and Combined Flame and Graphite Furnace models have a data coded
Air/Acetylene burner included.
Features:
Motorised 8 hollow cathode lamp (HCL) turret for automatic positioning and
optimisation of each HCL by the software;
Automatic switchover of spectral bandwidths (5 choices);
Advanced Graphite Furnace Horizontally heated platform graphite tube;
Computer controlled heating programmes which enables the user to select the best
programme for the analysis;
The optical temperature during atomisation stage ensures rapid heating and
analysis.
Technical Data
C10-9660 + C10-9673
Optical System
Wavelength range: 185nm ‒ 910nm
Monochromator: Czerny-Turner configuration
Spectral bandwidth: 0.1, 0.2, 0.4, 1.0 & 2.0nm
Wavelength Accuracy: ±0.15nm
Wavelength Reproducibility: < 0.05nm
Resolution: <30% or 0.2nm ±0.02nm
Baseline stability: 0.005A/30min
Flame Analysis
Sensitivity (Cu): 2.0μg/ml absorption > 0.2Abs
Detection limit (Cu): Cu < 0.0004μg/ml (flame);
Cd < 0.4 x 10 -12 g (graphite furnace)
Repeatability: Cu < 0.7% (Air-Acetylene analysis)
Ba < 1% (Nitrous Oxide analysis)
Cu < 2%, Cd < 2% (Graphite furnace analysis)
Background correction: Deuterium Arc, Self Reversal
Character Concentration: <0.02 μg/ml/1% (Cu); < 0.15mg/ml/1%
(Nitrous Oxide-Acetylene, Ba)
Character value: Cu < 0.3 x 10 -10 g, Cd < 0.3 x 10 -12 g
Burner head: Titanium alloy burner
Nebuliser: High efficiency glass nebuliser
Atomiser Chamber: Corrosion-resistant material
C10-9660
Position Adjustment: Automatic changeover of flame and furnace
Automatic setting of optimum height for flame
burner
Safety: Burner identification, Flame Sensor, Gas leak sensor,
Low gas pressure sensor, Power loss protection,
Gas loss flame protection
Mainframe
Light source: 8 HCL lamp turrets with 2 lamps simultaneously lit
Power supply: 110V/60Hz or 220V/50Hz, Three phase