Conductivity Meters
CONDUCTIVITY/TDS/SALINITY/TEMPERATURE METER, hand held
Waterproof, portable conductivity meter with TDS, resistivity and three salinity
scales. This meter automatically recognises the probe type (two or four ring)
and allows the user to adjust the nominal cell constant.
Capable of performing all three stages USP <645> method required for EC mea-
surement of ultrapure water. Choose from seven memorised standards and
obtain up to a five-point conductivity calibration. For salinity (% range), C30
5820 standard allows users to make a one-point calibration.
EC and TDS measurements are fully customisable and include: cell constant
selection, selection of linear or natural water (non-linear) or no temperature
compensation (for actual conductivity reading), configurable temperature com-
pensation coefficient range, choice of reference temperatures and a selectable
TDS factor. Features data logging transferrable to a PC with the supplied USB
cable and software.
Technical Data
Ranges: Cond. 0 to 400mS/cm (shows values up to 1000mS/cm),
Actual cond. 1000mS/cm; 0.001-9.999µS/cm; 10.00-99.99µS/cm;
100.0-999.9µS/cm; 1.000-9.999mS/cm; 10.00
99.99mS/cm; 100.0 to 1000.0mS/cm (auto ranging)
TDS 0.00-99.99mg/l (ppm); 100.0-999.9mg/l (ppm);
1.000-9.999g/l (ppt); 10.00-99.99g/l (ppt);
100.0-400.0g/l (ppt) (auto ranging)
Resistivity 1.0-99.9Ωcm; 100-999Ωcm; 1.00-9.99KΩcm;
10.0-99.9KΩcm; 100-999KΩcm; 1.00-9.99MΩcm;
10.0-100.0MΩcm (auto ranging)
NaCl 0.0-400.0%
Temp. -20.0-120.0ºC
Resolution: Cond. 0.001µS/cm; 0.01µS/cm; 0.1µS/cm; 0.001mS/cm;
0.01mS/cm; 0.1mS/cm
TDS 0.01mg/l (ppm); 0.1mg/l (ppm); 0.001g/l (ppt);
0.01g/l (ppt); 0.1g/l (ppt)
Resistivity: 0.1Ωcm; 1Ωcm; 0.01K Ωcm; 0.1KΩcm; 1KΩcm;
0.01MΩcm; 0.1MΩcm
Salinity: 0.1%; 0.01
Temp. 0.1ºC
Accuracy: Cond. ±1% of reading ±(0.01µS/cm or
1 digit, whichever is greater)
TDS ±1% of reading ±(0.05mg/l (ppm) or
1 digit, whichever is greater)
Resistivity: ±1% of reading (±10Ω or 1 digit, whichever is greater) C30-5850
NaCl ±1% of reading
Temp. ±0.2ºC (excluding probe error)
Calibration: Cond. automatic up to five points with seven memorised
standards (0.00µS/cm, 84.0µS/cm, 1.413mS/cm,
5.00mS/cm, 12.88mS/cm, 80.0mS/cm, 111.8mS/cm)
NaCl: 1 point with calibration buffer
Temp: 1 or 2 points
Cell Constant Setup: 0.010 to 10.000
Temp. Compensation: NoTC, linear (-20.0 to 120.0°C), non-linear (0 to 36°C)
ISO/DIS 7888 std
Reference Temp.: 15°C, 20°C and 25°C
ATC: 0.00 to 10.00 %/°C
TDS factor: 0.40 to 1.00
Data logging: log-on demand: 400 samples; lot logging: 5, 10, 30sec,
1, 2, 5, 10, 15, 30, 60, 120, 180min (max 1000 samples)
Memorised profiles: up to 10
Measurement modes: auto range, auto end, lock and fixed range
Auto-off: user selectable: 5, 10, 30, 60min, disabled
Supplied complete with platinum four ring conductivity/TDS probe, 250ml
1413µS/cm conductivity calibration solution, 250ml 111800µS/cm EC value
@25°C solution, rugged carrying case, 2 x 100ml plastic beakers, PC software,
micro-USB cable, batteries, instruction manual and quality certificate.
C30-5850 Conductivity/TDS/NaCl/Temperature Meter