Page 172 - Eagle Scientific Ltd - Catalogue
P. 172

ES
        Flame Photometer



         FLAME PHOTOMETERS

         FLAME PHOTOMETERS, JENWAY, PFP7/PFP7C
         Single beam emission flame photometers for routine determination of Na, K, Ca,
         Ba and Li, in industrial and clinical applications. Features include electronic flame
         failure detection, fine and coarse sensitivity controls, 3½ digit (12.5mm) LED and
         recorder output. Operates on propane, butane, natural gas or L.P.G. supplies. The
         model C40­6055 (PFP7C) is designed specifically for use in clinical applications. The
         built-in lineariser circuitry enables readings of both sodium and potassium, at
         normal clinical serum concentrations, to be directly displayed in mmol/l. Serum
         samples need to be diluted 200:1 or 100:1 prior to presentation to the flame
         photometer.
         Technical Data
                                                 C40-6005                        C40-6055
         Ranges:                          0-199.9                            120-160mmol/l Na (linearised)
                                                                                           0-10.0mmol/l K


         Limits of detection:      Na/K <0.2ppm                Na/K <0.2ppm
                                                 Li <0.25ppm                    Li <0.25ppm
                                                 Ca <15ppm
                                                 Ba <30ppm
         Reproducibility:            1% CV for 20 consecutive samples using 10ppm
                                                 Na set to read 50
         Linearity:                       Better than 2% when concentration of 3ppm
                         Na/K and 5ppm Li are set to read 100
         Specificity:                     Interference from Na, K, Li equal in concentration
                         to test element will be <0.5%
         Recorder output:         Nominal 1.00V for a reading of 100.0
         Dimensions:                  420 x 360 x 300mm
         Weight:                          8kg
         Power supply:               90-125V or 190-250V a.c. single phase              C40-6005
         Gas supply:                    Propane, butane or natural gas, (specify
                                                 regulator when ordering)
                                                          2
         Air supply:                     Moisture and oil free, 6 litres/min at 1kg/cm
         Industrial version is supplied with Na, K, Li, Ba & Ca filters.

         Clinical version is supplied with Na, K and Li filters.
         Both models are supplied with connecting hoses & clips, compressor plug, drain
         trap, fuses, mains lead and combined instruction book and service manual.
         Requires but does not include a gas regulator and an air compressor.
         The instrument will only operate with the correct type of air compressor and gas
         regulator - please specify requirements at time of order.
         C40-6005  Flame Photometer, PFP7, industrial
         C40-6055  Flame Photometer, PFP7C, clinical




























         166
   167   168   169   170   171   172   173   174   175   176   177